Online automatic intrinsic camera calibration for surgical endoscopes using a Gaussian Sum Filter embedded in a Monocular SLAM pipeline.
Developed as an M.Sc. thesis project at Technische Universität München (TUM) in collaboration with SCOPIS GmbH (surgical navigation).
Surgical endoscopes present a uniquely difficult calibration scenario:
- Varying intrinsics — focus and zoom adjustments during a procedure change focal length in real time
- Lens distortion — wide-angle optics introduce significant radial distortion
- No calibration target — a checkerboard cannot be used inside a patient
This project estimates the full intrinsic parameter set [f, Cx, Cy, k1, k2] online, directly from a live endoscope image stream, without interrupting the surgical workflow.
A bank of Extended Kalman Filters runs in parallel. Each filter is initialised with a different hypothesis for focal length f and radial distortion coefficients k1, k2, spanning the plausible range for the endoscope optics.
FilterBank
├── EKF₁ [f = f_min, k1 = k1_start, k2 = k2_start]
├── EKF₂ [f = f_min + Δf, ...]
│ ...
└── EKFₙ [f = f_max, k1 = k1_end, k2 = k2_end ]
Each filter maintains a full MonoSLAM state: camera pose (3-DoF translation + quaternion orientation) + tracked feature map + camera intrinsics. All filters receive the same image observations; filters whose intrinsic hypotheses conflict with the observed feature motion accumulate low likelihood and are eliminated.
After each frame a log-likelihood ratio is computed across the filter bank. The SPRT with thresholds A and B (derived from user-specified false-positive and false-negative error rates) decides when one hypothesis has accumulated sufficient evidence — pruning the bank until a single calibration estimate remains.
A = log₁₀((1 - β) / α) B = log₁₀(β / (1 - α))
α = type-I error rate β = type-II error rate
Patch-based template matching (OpenCV) detects and tracks corner features across consecutive frames. Matched correspondences feed the EKF measurement update for both the 3-D map and the intrinsic parameters simultaneously.
| Component | Dim | Description |
|---|---|---|
f |
1 | Focal length (pixels) |
Cx, Cy |
2 | Principal point |
k1, k2 |
2 | Radial distortion coefficients |
r_wc |
3 | Camera translation (world frame) |
q_wc |
4 | Camera orientation (unit quaternion) |
| Features | 3×N | 3-D map point positions |
Mono_Exp_EndoIm/
├── Mono_Exp.cpp # Entry point — frame loop & SPRT output
├── Monoslam.h / .cpp # Top-level SLAM coordinator
├── FilterBank.h / .cpp # GSF: pool of EKF instances
├── KalmanFilter.h / .cpp # Single EKF (predict + update)
├── MotionModel.h / .cpp # 6-DoF quaternion motion model
├── Camera.h / .cpp # Intrinsic parameter model
├── DataAssociator.h / .cpp # Feature matching & data association
├── MathUtil.h / .cpp # Quaternion / Jacobian utilities
├── cali_results.txt # Output log of the current run (720×576 sequence)
└── Calibration_results/ # Archived logs from an earlier 320×240 sequence
| Library | Purpose |
|---|---|
| OpenCV ≥ 2.4 | Image I/O, feature detection, patch matching |
| Eigen ≥ 3.2 | Matrix algebra for EKF state and covariance |
| C++11 compiler | MSVC 2013+ (original); GCC/Clang with minor porting |
Originally developed with Visual Studio on Windows. To build on Linux/macOS, replace sprintf_s with snprintf in Mono_Exp.cpp and use the following CMakeLists.txt:
cmake_minimum_required(VERSION 3.5)
project(MonoSLAMAutoCalibration)
find_package(OpenCV REQUIRED)
find_package(Eigen3 REQUIRED)
file(GLOB SOURCES "Mono_Exp_EndoIm/*.cpp")
add_executable(mono_slam ${SOURCES})
target_include_directories(mono_slam PRIVATE ${EIGEN3_INCLUDE_DIR})
target_link_libraries(mono_slam ${OpenCV_LIBS})mkdir build && cd build
cmake ..
make -j$(nproc)- Place the endoscope image sequence under
Mono_Exp_EndoIm/exp_scopis_im_2_720x576/asstep<N>.jpg - Set the start frame and range in
Mono_Exp.cpp:
int step_ = 344; // starting frame index
int lastIm = 900; // last frame- Provide camera pixel-size and resolution in
camdata/cam.txt - Run the binary — calibration estimates are appended to
cali_results.txt:
frame f (px) Cx (px) Cy (px) k1 k2 filters_alive
344 490.873 362.551 291.323 0.041420 0.009722 72
345 498.592 363.407 293.725 0.021161 0.003104 19
...
353 500.035 363.253 295.252 0.020013 0.003002 1
...
441 433.478 354.582 285.387 0.019108 0.002792 1
Each row has 17 columns: the 7 shown above, followed by the ±3σ bounds for
f, Cx, Cy, k1, k2 (two columns each). Note that k1 and k2 are in
1/mm² and 1/mm⁴ — the radial distortion is applied to a metric radius
obtained via the pixel pitch dx from camdata/cam.txt, so these are not the
dimensionless coefficients used by OpenCV.
The filter bank is built with 108 parallel hypotheses (18 focal lengths × 2 k1
× 3 k2, see FilterBank::initialize_filterbank). Pruning happens inside the
first processed frame, so the first logged row already shows 72 survivors.
Each ellipse is the EKF predicted search region for a map feature; the dot is the measured pixel position.
As the filter bank accumulates observations, ellipses shrink — reflecting reduced uncertainty in both feature positions and camera intrinsics.
| 🔴 Red ellipse | 🔵 Blue ellipse |
|---|---|
| High uncertainty — newly initialised or rarely observed | Low uncertainty — well-established, frequently matched |
If the video does not render, view the animation below or download the MP4.
| Step 344 — Initialisation | Step 350 — Early tracking | Step 360 — Building map |
|---|---|---|
| Large red search regions | Red/blue mix, camera moving | Blue features appearing |
![]() |
![]() |
![]() |
| Step 380 — Established tracking | Step 415 — Dense map | Step 441 — Converged |
|---|---|---|
| Mostly blue, ellipses tightening | Tight blue ellipses, few red | Small compact ellipses |
![]() |
![]() |
![]() |
The circular field of view and radial barrel distortion are characteristic of the wide-angle endoscope optics. The GSF estimates
f,Cx,Cy,k1,k2online from exactly this imagery — no calibration target required.
Focal length, principal point, and active filter count over the 720×576
sequence, plotted from Mono_Exp_EndoIm/cali_results.txt. Dashed lines mark the
offline SCOPIS calibration; the shaded band on f is the ±3σ the filter reports.
The bank is built with 108 hypotheses, drops to 72 within the first frame, and
reaches a single survivor at frame 353.
Tested on a 720×576 SCOPIS endoscope sequence, steps 344–441 (98 frames). The bank is pruned to a single surviving hypothesis at frame 353 — nine frames in — after which the run is effectively a single EKF refining that hypothesis.
Values below are the mean over the last 20 frames (422–441), compared against
the offline SCOPIS calibration that the code carries as its reference baseline
(Kalman.cpp, hi_cartesian_scopis):
| Parameter | GSF estimate | SCOPIS reference | Difference |
|---|---|---|---|
Focal length f |
433.1 px | 440 px | −1.6 % |
Principal point Cx |
354.6 px | 340 px | +14.6 px |
Principal point Cy |
285.4 px | 277 px | +8.4 px |
Radial distortion k1 |
0.01911 1/mm² | 0.0101 1/mm² | +89 % |
Radial distortion k2 |
0.002792 1/mm⁴ | 0.00146 1/mm⁴ | +91 % |
Focal length lands within 2 % of the reference. The principal point settles to a stable value 8–15 px away from it, and the two distortion coefficients come out roughly twice the reference — at the edge of the endoscope's circular field of view (r ≈ 350 px) that is a radial magnification of 1.95 against the reference's 1.50.
The ±3σ interval reported in cali_results.txt narrows to ±4.3 px on f and
±0.5 px on Cx by frame 441, which does not bracket the reference for any of
the five parameters. The intrinsic states carry no process noise, so once the
bank has collapsed their covariance can only shrink; the reported interval
should be read as filter self-consistency, not as accuracy.
This project is a direct implementation of the Gaussian Sum Filter for online camera self-calibration described in:
J. Civera, A. J. Davison, and J. M. M. Montiel, "Inverse Depth Parametrization for Monocular SLAM," IEEE Transactions on Robotics, vol. 24, no. 5, pp. 932–945, Oct. 2008. DOI: 10.1109/TRO.2008.2003276 · PDF
The paper introduces two key contributions implemented here:
- Inverse depth parametrization for monocular feature initialisation, improving EKF linearisation accuracy over Cartesian coordinates
- Gaussian Sum Filter (Section IV) — a bank of EKFs with different camera intrinsic hypotheses, pruned via a Sequential Probability Ratio Test, enabling online self-calibration without a calibration target
The application to surgical endoscope cameras (SCOPIS system, 720×576) extends the method to a medical imaging domain where intrinsics change actively during use.
Additional references:
- Davison, A. J. et al. — MonoSLAM: Real-Time Single Camera SLAM, IEEE TPAMI, vol. 29, no. 6, 2007
- Civera, J., Grasa, O. G., Davison, A. J., Montiel, J. M. M. — 1-Point RANSAC for EKF Filtering, IROS 2009
Shulin Gao — M.Sc. Computational Science & Engineering, TU Munich
linkedin.com/in/shulin-gao · github.com/slgao